Type | Adapter I |
Model 0 | Model I | Model II | Model IV | |||
Max. working volume (ml) | 50 | 100 | 200 | 300 | 250 | 500 |
Max. working pressure (bar) | 100 | 100 | 100 | 100 | 200 | 100 |
Heating mantle | none | 10 S | 20 S | 30 S | 30 S | 50 S |
Sampling possible | no | no | yes | yes | ||
Stirrer connection possible | no | no | yes | yes | ||
Temp. measurement in autoclave possible | no | no | yes | yes |
Due to the shape of the screw thread, cold setting of cylinder and head is possible when screwing these on. This usually leads to them sticking when opening and the thread breaking as a result.The conducted reaction necessitates coating the thread with a release agent such as PTFE spray (Art. No. K007.1).Warning: Do not use PTFE screw thread tape!
When working with silver gaskets, it is essential to tighten the screw threads very tightly to achieve the necessary seat. We therefore recommend annealing the silver gaskets at +850 °C before use - the gaskets remain softer and more pliable, even when cool, and seal more efficiently.